IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 1 SUBSIDIARY LEGISLATION 427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS 1st January, 2003 LEGAL NOTICE 61 of 2002, as amended by Legal Notices 30 of 2004, 21 of 2010, 371 of 2012 and 318 of
(2)of the Directive and published in the OJ, as set out in Schedule XI. Applicability. 4.1. The provisions of these regulations are applicable to all imported in vitro diagnostic medical devices as well as those manufactured for the local market. 4.1.1. No person may manufacture, import, place on the market or put into service any in vitro diagnostic medical devices which do not comply with the provisions of these regulations. 4.2. Subject to regulation 4.3, the provisions of these regulations shall not apply to in vitro diagnostic medical devices intended exclusively for export. 4.3. Subject to regulation 11, regulation 4.2 shall not apply if the CE marking or any inscription liable to be confused therewith is affixed to the in vitro diagnostic medical devices. General requirements. Amended by: L.N. 30 of 2004. 5.1. All necessary steps shall be taken to ensure that devices are placed on the market and, or put into service only if they comply with the requirements laid down in these regulations when duly supplied and properly installed, maintained and used in accordance with their intended purpose. 5.1.1. This regulation applies also to devices made available for performance evaluation. 5.2. Devices must meet the essential requirements set out in Annex I of the Directive, which is set out in Schedule I, which apply to them, taking account of the intended purpose of the devices concerned. 5.3. A device shall be treated as complying with an essential requirement if as respects that requirement, it complies with the relevant national standard, unless there are reasonable indications that the device does not comply with that requirement. 5.4. Without prejudice to regulation 5.3, a device specified in List A or B of Annex II of the Directive, set out in Schedule II, shall be treated as complying with the relevant essential requirements if the device is designed and manufactured in conformity with the common technical specifications drawn up for the device, unless there are reasonable indications that the device does not comply with those requirements. 5.4.1. A device to which regulation 5.4 relates shall be treated as having been manufactured in accordance with the common technical specifications drawn up for the device if, for justifiable reasons, it is not so manufactured but is manufactured in accordance with technical specifications that are at least equivalent to such common technical specifications, unless there are reasonable indications that the device does not comply with those requirements. 5.4.1.1. The technical specifications set out in IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 Schedule XI are adopted as common technical specifications for devices in list A of Schedule II. 5.5. No person shall place on the market any device to which these regulations apply unless the requirements of regulation 5.2 have been complied with in relation to it. 6.1. The placing on the market or putting into service under the conditions specified by the manufacturer of devices referred to in regulation 3, which comply with these regulations and bear the CE marking provided for in regulation 11, indicating that they have undergone conformity assessment in accordance with regulation 7 shall not be prohibited, restricted or impeded. Free movement. 6.2. No obstacle shall be created to devices intended for performance evaluation being made available for that purpose to the laboratories or other institutions listed in the statement referred to in Annex VIII of the Directive, which is set out in Schedule VIII, if they meet the conditions laid down in regulation 7.4 and Annex VIII of the Directive, which is set out in Schedule VIII. 6.3. At trade fairs, exhibitions, demonstrations, scientific or technical gatherings, etc., no obstacle shall be created to the showing of devices which do not conform to these regulations, provided that such devices are not used on specimens taken from the participants and that a visible sign clearly indicates that such devices cannot be marketed or put into service until they have been made to comply. 6.4. The information to be supplied pursuant to Annex I, part B, section 8 of the Directive, which is set out in Schedule I, is required to be in either the Maltese or English language, when the device reaches the final user. 6.4.1. Provided that safe and correct use of the device is ensured, the information referred to in regulation 6.4 may be authorised to be in one or more other official Community language(s). 6.4.1.1. In the application of this provision, account shall be taken of the principle of proportionality and, in particular: - whether the information can be supplied by harmonised symbols or recognised codes or other measures; - the type of user anticipated for the device. 7.1. For all devices other than those covered by Annex II of the Directive, which i s s et out in S c hed u le I I , a nd d ev i ces f or performance evaluation, the manufacturer shall, in order to affix the CE marking, follow the procedure referred to in Annex III of the Directive, which is set out in Schedule III, and draw up the EC declaration of conformity required before placing the devices on the market. 7.1.1. For all devices for self-testing other than those covered by Annex II of the Directive, which is set Conformity assessment procedures. 7 8 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES out in Schedule II, and medical devices for performance evaluation, the manufacturer shall, prior to the drawing up of the aforementioned declaration of conformity, fulfil the supplementary requirements set out in Annex III, point 6 of the Directive, which is set out in Schedule III. Instead of applying this procedure, the manufacturer may follow the procedure referred to in regulations 7.2 and 7.3. 7.2. For all devices referred to in List A in Annex II of the Directive, which is set out in Schedule II, other than those intended for performance evaluation, the manufacturer shall, in order to affix the CE marking either: 7.2.1. follow the procedure relating to the EC declaration of conformity set out in Annex IV (full quality assurance) of the Directive, which is set out in Schedule IV, or 7.2.2. follow the procedure relating to EC typeexamination set out in Annex V of the Directive, which is set out in Schedule V, coupled with the procedure relating to the EC declaration of conformity set out in Annex VII (production quality assurance) of the Directive, which is set out in Schedule VII. 7.3. For all devices referred to in List B in Annex II of the Directive, which is set out in Schedule II, other than those intended for performance evaluation, the manufacturer shall for the purposes of affixing the CE marking, follow either: 7.3.1. the procedure relating to the EC declaration of conformity set out in Annex IV (full quality assurance) of the Directive, which is set out in Schedule IV, or 7.3.2. the procedure relating to EC type-examination set out in Annex V of the Directive, which is set out in Schedule V, coupled with: 7.3.2.1. 7.3.2.2. the procedure relating to EC verification set out in Annex VI, which is set out in Schedule VI, or the procedure relating to the EC declaration of conformity set out in Annex VII (production quality assurance), which is set out in Schedule VII. 7.4. In the case of devices for performance evaluation, the manufacturer shall follow the procedure referred to in Annex VIII of the Directive, which is set out in Schedule VIII, and draw up the statement set out in that Schedule before such devices are made available. 7.4.1. This provision does not affect national regulations relating to the ethical aspects of carrying out IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 performance evaluation studies using tissues or substances of human origin. 7.5. During the conformity assessment procedure for a device, the manufacturer and, if involved, the notified body shall take account of the results of any assessment and verification operations which, where appropriate, have been carried out in accordance with these regulations at an intermediate state of manufacture. 7.6. The manufacturer may instruct his authorised representative to initiate the procedures provided for in Annexes III, V, VI and VIII of the Directive, which are set out in Schedules III, V, VI and VIII. 7.7. The manufacturer must keep the declaration of conformity, the technical documentation referred to in Annexes III to VIII of the Directive, which are set out in Schedules III and VIII, as well as the decisions, reports and certificates, established by notified bodies, and make it available to the national authorities for inspection purposes for a period ending five years after the last product has been manufactured. Where the manufacturer is not established in Malta, the obligation to make the aforementioned documentation available on request applies to his authorised representative. 7.8. Where the conformity assessment procedure involves intervention of a notified body, the manufacturer, or his authorised representative, may apply to a body of his choice within the framework of tasks for which the body has been notified. 7.9. The notified body may require, where duly justified, any information or data, which is necessary for establishing and maintaining the attestation of conformity in view of the chosen procedure. 7.10. Decisions taken by the notified bodies in accordance with Annexes III, IV, and V of the Directive, which are set out in Schedules III, IV, and V, shall be valid for a maximum of five years and may be extended on application, made at a time agreed in the contract signed by both parties, for further periods of up to five years. 7.11. The records and correspondence relating to the procedures referred to in regulations 7.1 to 7.4 shall be in an official language. 7.12. By way of derogation from regulations 7.1 to 7.4, the competent authorities may authorise, on duly justified request, the placing on the market and putting into service, of individual medical devices for which the procedures referred to in regulations 7.1 to 7.4 have not been carried out and the use of which is in the interest of protection of health. 7.13. The provisions of regulation 7 shall apply accordingly to any natural or legal person who manufactures devices covered by these regulations and, without placing them on the market, puts them into service and uses them in the context of his professional activity. 9 10 [ S.L.427.16 Withdrawal and prohibition from market. Amended by: L.N. 318 of 2020. IN VITRO DIAGNOSTIC MEDICAL DEVICES 8.1. Where it is ascertained that the devices referred to in regulation 6.1, when correctly installed, maintained and used in accordance with their intended purpose, may compromise the health and, or safety of patients, users or, where applicable other persons, or the safety of property, the Medicines Authority shall take all appropriate interim measures to withdraw such devices from the market, or prohibit or restrict their being placed on the market or put into service. 8.1.1. The Medicines Authority shall immediately inform Commission of any such measures, indicating the reasons for the decision and in particular, whether non-compliance is due to: 8.1.1.1. 8.1.1.2. 8.1.1.3. failure to meet the essential requirements referred to in regulation 5; incorrect application of the standards referred to in Article 5 of the Directive, insofar as it is claimed that the standards have been applied; shortconungs in the standards themselves. 8.2. Where a non-complying medical device referred to in regulation 3 bears the CE marking, the Medicines Authority shall take appropriate action against whomsoever has affixed the marking or drawn the declaration and shall so inform the Commission and the other Member States thereof. 8.3. Any decision taken pursuant to these regulations which restricts the placing on the market and putting into service of devices shall state the exact grounds on which it is based. 8.3.1. Such a decision shall be notified as soon as possible to the party concerned, who shall at the same time be informed of the legal remedies available to him under the laws in force in Malta and of the time limits to which such remedies are subject. 8.4. Where the Medicines Authority considers, in relation to a given product or group of products, that, in order to ensure protection of health and safety and, or to ensure that public health requirements are observed pursuant to Article 36 of the Treaty, the availability of such products should be prohibited, restricted or made subject to particular requirements, necessary and justified transitional measures may be taken. 8.5. The Medicines Authority shall immediately inform the Commission of any measures taken in pursuance of regulations 8.1.1, 8.2 and 8.4. 8.6. In the event of a decision as referred to in regulation 8.2, the manufacturer or his authorised representative shall have an opportunity to put forward his point of view in advance, unless such consultation is not possible because of the urgency of the measure to be taken as justified in particular by public health requirements. IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 9.1. Any manufacturer who places devices on the Maltese market under his own name shall notify the Medicines Authority: 9.1.1. of the address of the registered place of business, 9.1.2. in the case of an authorised representative, sufficient evidence that he is the authorised representative of the manufacturer; 9.1.3. of information relating to the reagents, reagent products and calibration and control materials in terms of common technological characteristics and, or analytes and of any significant change thereto including discontinuation of placing on the market; for other devices, the appropriate indications; 9.1.4. in the case of devices covered by Annex II of the Directive, which is set out in Schedule II, and of devices for self-testing, of all data allowing for identification of such devices, the analytical and, where appropriate, diagnostic parameters as referred to in Annex I, part A, section 3 of the Directive, which is set out in Schedule I, the outcome of performance evaluation pursuant to Annex VIII of the Directive, which is set out in Schedule VIII, certificates and any significant change thereto, including discontinuation of placing on the market. 9.2. For devices covered by Annex II of the Directive, which is set out in Sched ule I I, and for devices for self-t esting, the Medicines Authority may request to be informed of the data allowing identification together with the label and the instructions for use when such medical devices are placed on the market and, or put into service within their territory. 9.2.1. The measures given in regulation 9.2 cannot constitute a precondition for the placing on the market and, or putting into service of medical devices which are in conformity with these regulations. 9.3. Where a manufacturer who places medical devices on the market under his own name does not have a registered place of business in Malta, he shall designate an authorised representative. The authorized representative shall notify the Medicines Authority of all paiticulars as referred to in regulation 9.1. 9.4. The notification referred to in regulation 9.1 shall also include any new device. In addition, where, in the context of such notification, a device notified, bearing the CE marking, is a "new product", the manufacturer shall indicate this fact on his notification. if: For the purposes of regulation 9, a medical device is "new" 9.4.1. there has been no such device continuously available on the Cornmunity market during the previous three years for the relevant analyte or 11 Registration of manufacturers and devices. Amended by: L.N. 318 of 2020. 12 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES other parameter; 9.4.2. the procedure involves analytical technology not continuously used in connection with a given analyte or other parameter on the Community market during the previous three years. 9.5. The notifications referred to in regulations 9.1 and 9.3 are to be registered immediately in the databank described in Article 12 of the Directive. The procedures for implementing this regulation and in particular those referring to the notification and the concept of significant change shall be adopted in accordance with the procedure referred to in Article 7 of the Directive. 9.6. Transitionally, pending the establishment of a European Databank accessible to the competent authorities of the Member States and containing the data relating to all medical devices available on the territory of the Comrnunity, the manufacturer shall give such notification to the competent authorities of each Member State concerned by the placing on the market. Vigilance procedure. Amended by: L.N. 318 of 2020. 10.1. The Medicines Authority shall take the necessary steps to ensure that any information brought to his knowledge, in accordance with the provisions of these regulations, regarding the incidents mentioned below involving devices bearing the CE marking is recorded and evaluated centrally: 10.1.1. any malfunction, failure or deterioration in the characteristics and, or performance of a device, as well as any inadequacy in the labelling or the instructions for use which, directly or indirectly, might lead to or might have led to the death of a patient, or user or of other persons or to a serious deterioration in their state of health; 10.1.2. any technical or medical reason in relation to the characteristics or performance of a device for the reasons referred to in regulation 10.1.1, leading to systematic recall of medical devices of the same type by the manufacturer. 10.2. Where medical practitioners, medical institutions or the organizers of external quality assessment schemes are required to inform the competent authorities of any incidents referred to in regulation 10.1, necessary steps shall be taken to ensure that the manufacturer of the device concerned, or his authorized representative, is also informed on the incident. 10.3. After carrying out an assessment, if possible together with the manufacturer, the Medicines Authority shall, without prejudice to Article 8, immediately inform the Commission and the other Member States of the incidents referred to in regulation 10.1 for which appropriate measures, including possible withdrawal, have been taken or are contemplated. 10.4. Where, in the context of notification referred to in regulation 9, a device notified, bearing the CE marking, is a "new" product, the manufacturer shall indicate this fact on his notification. The Medicines Authority so notified may at any time IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 13 within the following two years and on justified grounds, require the manufacturer to submit a report relating to the experience gained with the device subsequent to its being placed on the market. 10.5. The Medicines Authority of the Malta Standards Authority shall immediately inform the Commission of any measures taken in pursuance of regulation 10.3 of these regulations. 11.1. The CE marking consists of the initials "CE" in accordance with the model in Annex X of the Directive, which is set out in Schedule X. CE marking. 11.2. Devices, other than medical devices for performance evaluation, considered to meet the essential requirements referred to in regulation 5, must bear the CE marking of conformity when they are placed on the market. 11.3. The CE marking of conformity, as shown in Annex X of the Directive, which is set out in Schedule X, must appear in a visible, legible and indelible form on the device, where practicable and appropriate, and on the instructions for use. The CE marking of conformity must also appear on the sales packaging. 11.3.1. The CE marking shall be accompanied by the identification number of the notified body responsible for implementation of the procedures set out in Annexes III, IV, VI and VII of the Directive, which are set out in Schedules III, IV, VI, and VII. 11.4. No person shall affix on any device, to which these regulations apply, any markings or inscriptions which are likely to mislead third parties with regard to the meaning or the graphics of the CE marking. 11.5. Without prejudice to regulation 11.4, nothing in these regulations shall prohibit the affixing on device, to the packaging or to the instruction leaflet accompanying the device, provided that the visibility and legibility of the CE marking are not thereby reduced. 11.6. Without prejudice to regulation 8: 11.6.1. where it is established that the CE marking has been wrongly affixed, the manufacturer or his authorized representative shall be obliged to end the infringement under conditions laid down by the Medicines Authority; 11.6.2. where non-compliance continues, the Medicines Authority shall take all appropriate measures to restrict or prohibit the placing on the market of the product in question or to ensure that it is withdrawn from the market, in accordance with the procedure laid down in regulation 8. 11.7. The provisions stated in regulation 11.6 shall also apply where the CE marking has been affixed in accordance with the procedures in these regulations, but inappropriately, on products that are not covered by these regulations. Amended by: L.N. 318 of 2020. 14 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES 11.8. Subject to regulation 11.9, where the devices are subject to other directives, concerning other aspects which also provide for the affixing of the CE marking, the latter shall indicate that the devices also fulfil the provisions of the other directives. 11.9. Where one or more of the other directives referred to in regulation 11.8 allow the manufacturer, during a transitional period, to choose which arrangements to apply, the CE marking shall indicate that the devices fulfil the provisions only of those directives applied by the manufacturer. In case, the particulars of these directives, as published in the Official Journal of the European Communities, must be given in the documents, notices or instructions required by the directives and accompanying such devices. Confidentiality. 12.1. Without prejudice to national law and practice on medical secrecy, it shall be ensured that all the parties involved in the application of these regulations are bound to observe confidentiality with regard to information obtained in carrying out their tasks. This does not affect the obligations of the persons concerned to provide information under criminal law. IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 15 SCHEDULE I Based on Annex I of the Directive ESSENTIAL REQUIREMENTS A. GENERAL REQUIREMENTS 1. The devices must be designed and manufactured in such a way that, when used under the conditions and for the purposes intended, they will not compromise, direcdy or indirectly, the clinical condition or the safety of the patients, the safety or health of users or, where applicable, other persons, or the safety of property. Any risks, which may be associated with their use, must be acceptable when weighed against the benefits to the patient and be compatible with a high level of protection of health and safety. 2. The solutions adopted by the manufacturer for the design and construction of the devices must conform to safety principles, taking account of the generally acknowledged state of the art. In selecting the most appropriate solutions, the manufacturer must apply the following principles in the following order: - eliminate or reduce risks as far as possible (inherently safe design and construction), - where appropriate take adequate protection measures in relation to risks that cannot be eliminated, - inform users of the residual risks due to any shortcomings of the protection measures adopted. 3. The devices must be designed and manufactured in such a way that they are suitable for the purposes referred to in regulation 3.1.3, as specified by the manufacturer, taking account of the generally acknowledged state of the art. They must achieve the performances, in particular, where appropriate, in terms of analytical sensitivity, diagnostic sensitivity, analytical specificity, diagnostic specificity, accuracy, repeatability, reproducibility, including control of known relevant interference, and limits of detection, stated by the manufacturer. The traceability of values assigned to calibrators and, or control materials must be assured through available reference measurement procedures and, or available reference materials of a higher order 4. The characteristics and performances referred to in points 1 and 3 of this section must not be adversely affected to such a degree that the health or the safety of the patient or the user and, where applicable, of other persons, are compromised during the lifetime of the device as indicated by the manufacturer, when the device is subjected to the stresses which can occur during normal conditions of use. When no lifetime is stated, the same applies for the lifetime reasonably to be expected of a device of that kind, having regard to the intended purpose and the anticipated use of the device. 5. The devices must be designed, manufactured and packed in such a way that their characteristics and performances during their intended use will not be adversely affected under storage and transport conditions (temperature, humidity, etc.) taking account of the instructions and information provided by the manufacturer. B. DESIGN AND MANUFACTURING REQUIREMENTS 1. Chemical and physical properties 1.1. The devices must be designed and manufactured in such a way as to achieve the characteristics and performances referred to in section A on the "General requirements". Particular attention must be paid to the possibility of impairment of 16 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES analytical performance due to incompatibility between the materials used and the specimens (such as biological tissues, cells, body fluids and micro-organisms) intended to be used with the device, taking account of its intended purpose. 1.2. The devices must be designed, manufactured and packed in such a way as to reduce as far as possible the risk posed by product leakage, contaminants and residues to the persons involved in the transport, storage and use of the medical devices, taking account of the intended purpose of the products. 2. Infection and microbial contamination 2.1. The devices and their manufacturing processes must be designed in such a way as to eliminate or reduce as far as possible the risk of infection to the user or other persons. The design must allow easy handling and, where necessary, reduce as far as possible contamination of, and leakage from, the device during use and, in the case of specimen receptacles, the risk of contamination of the specimen. The manufacturing processes must be appropriate for these purposes. 2.2. Where a device incorporates biological substances, the risks of infection must be reduced as far as possible by selecting appropriate donors and appropriate substances and by using appropriate, validated inactivation, conservation, test and control procedures. 2.3. Devices labelled either as "STERILE" or as having a special microbiological state must be designed, manufactured and packed in an appropriate pack, according to procedures suitable for ensuring that they remain in the appropriate microbiological state indicated on the label when placed on the market, under the storage and transport conditions specified by the manufacturer, until the protective packaging is damaged or opened. 2.4. Devices labelled either as "STERILE" or as having a special microbiological state must have been processed by an appropriate validated method. 2.5. Packaging systems for devices other than those referred to in point 2.3 of this section must keep the product without deterioration at the level of cleanliness indicated by the manufacturer and, if the devices are to be sterilised prior to use, reduce as far as possible the risk of microbial contamination. Steps must be taken to reduce as far as possible microbial contamination du rin g sel ectio n an d h andli ng of raw m aterials, m anufactu re, storage and distribution where the performance of the device can be adversely affected by such contamination. 2.6. Devices intended to be sterilised must be manufactured in appropriately controlled (e.g. environmental) conditions. 2.7. Packaging systems for non-sterile devices must keep the product without deterioration at the level of cleanliness stipulated and, if the devices are to be sterilised prior to use, minimise the risk of microbial contamination; the packaging system must be suitable taking account of the method of sterilisation indicated by the manufacturer. 3. Manufacturing and environmental properties 3.1. If the device is intended for use in combination with other devices or equipment, the whole combination, including the connection system, must be safe and must not impair the specified performances of the devices. Any restrictions on use must be indicated on the label and, or in the instructions for use. 3.2. Devices must be designed and manufactured in such a way as to reduce as far as possible the risks linked to their use in conjunction with materials, substances IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 17 and gases with which they may come into contact during normal conditions of use. 3.3. Devices must be designed and manufactured in such a way as to remove or reduce as far as possible: - the risk of injury linked to their physical features (in particular aspects of volume x pressure, dimension and, where appropriate, ergonomic features), - risks linked to reasonably foreseeable external influences, such as magnetic fields, external electrical effects, electrostatic discharge, pressure, humidity, temperature or variations in pressure or acceleration or accidental penetration of substances into the medical device. Devices must be designed and manufactured in such a way as to provide an adequate level of intrinsic immunity of electromagnetic disturbance to enable them to operate as intended. 3.4. Devices must be designed and manufactured in such a way as to reduce as far as possible the risks of fire or explosion during normal use and in single fault condition. Particular attention must be paid to devices whose intended use includes exposure to or use in association with flammable substances or substances which could cause combustion. 3.5. Devices must be designed and manufactured in such a way as to facilitate the management of safe waste disposal. 3.6. The measuring, monitoring or display scale (including colour change and other visual indicators) must be designed and manufactured in line with ergonomic principles, taking account of the intended purpose of the device. 4. Devices which are instruments or apparatus with a measuring function 4.1. Devices which are instruments or apparatus having a primary analytical measuring function must be designed and manufactured in such a way as to provide adequate stability and accuracy of measurement within appropriate accuracy limits, taking into account the intended purpose of the medical device and of available and appropriate reference measurement procedures and materials. The accuracy limits have to be specified by the manufacturer. 4.2. When values are expressed numerically, they must be given in legal units conforming to the provisions of Council Directive 80/181/EEC of 20 December 1979 on the approximation of the laws of the Member States relating to units of measurement. 5. Protection against radiation 5.1. Devices shall be designed, manufactured and packaged in such a way that exposure of users and other persons to the emitted radiation is minimised. 5.2. When devices are intended to emit potentially hazardous, visible and, or invisible radiation, they must as far as possible be: - designed and manufactured in such a way as to ensure that the characteristics and the quantity of radiation emitted can be controlled and, or adjusted, - fitted with visual displays and, or audible warnings of such emissions. 5.3. The operating instructions for medical devices emitting radiation must give detailed information as to the nature of the emitted radiation, means of protecting the user, and on ways of avoiding misuse and of eliminating the risks inherent in installation. 18 6. [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES Requirements for devices connected to or equipped with an energy source 6.1. Devices incorporating electronic programmable systems, including software, must be designed to ensure the repeatability, reliability and performance of these systems according to the intended use. 6.2. Devices must be designed and manufactured in such a way as to minimise the risks of creating electromagnetic perturbation which could impair the operation of other devices or equipment in the usual environment. 6.3. Devices must be designed and manufactured in such a way as to avoid, as far as possible, the risk of accidental electric shocks during normal use and in single fault condition, provided the devices are installed and maintained correctly. 6.4. Protection against mechanical and thermal risks 6.4.1. Devices must be designed and manufactured in such a way as to protect the user against mechanical risks. Devices must be sufficiently stable under the foreseen operating conditions. They must be suitable to withstand stresses inherent in the foreseen working environment, and to retain this resistance during the expected life of the devices, subject to any inspection and maintenance requirements as indicated by the manufacturer. Where there are risks due to the presence of moving parts, risks due to break-up or detachment, or leakage of substances, then appropriate protection means must be incorporated. Any guards or other means included with the device to provide protection, in particular against moving parts, must be secure and must not interfere with access for the normal operation of the medical device, or restrict routine maintenance of the device as intended by the manufacturer. 7. 6.4.2. Devices must be designed and manufactured in such a way as to reduce to the lowest possible level the risks arising from vibration generated by the devices, taking account of technical progress and of the means available for limiting vibrations, particularly at source, unless the vibrations are part of the specified performance. 6.4.3. Devices must be designed and manufactured in such a way as to reduce as far as possible the risks arising from the noise emitted, taking account of technical progress and of the means available to reduce noise, particularly at source, unless the noise emitted is part of the specified performance. 6.4.4. Terminals and connectors to electricity, gas or hydraulic and pneumatic energy supplies which the user has to handle must be designed and manufactured in such a way as to minimise all possible risks. 6.4.5. Accessible parts of the devices (excluding the parts of areas intended to supply heat or reach given temperatures) and their surroundings must not attain potentially dangerous temperatures under normal use. Requirements for devices for self-testing Devices for self-testing must be designed and manufactured in such a way that they perform appropriately for their intended purpose taking into account the skills and the means available to users and the influence resulting from variation that can reasonably be anticipated in users’ technique and environment. The information IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 19 and instructions provided by the manufacturer should be easily understood and applied by the user. 7.1. Devices for self-testing must be designed and manufactured in such a way as to: - ensure that the device is easy to use by the intended lay user at all stages of the procedure, and - reduce as far as practicable the risk of user error in the handling of the device and in the interpretation of the results. 7.2. Devices for self-testing must, where reasonably possible, include user control, i.e. a procedure by which the user can verify that, at the time of use, the product will perform as intended. 8. Information supplied by a manufacturer 8.1. Each device must be accompanied by the information needed to use it safely and properly, taking account of the training and knowledge of the potential users, and to identify the manufacturer. This information comprises the data on the label and in the instructions for use. As far as practicable and appropriate, the information needed to use the device safely and properly must be set out on the device itself and, or, where appropriate, on the sales packaging. If individual full labelling of each unit is not practicable, the information must be set out on the packaging and, or in the instructions for use supplied with one or more devices. Instructions for use must accompany or be included in the packaging of one or more devices. In duly justified and exceptional cases no such instructions for use are needed for a device if it can be used properly and safely without them. The decision whether to translate the instructions for use and the label into one or more languages of the European Union shall be left to the Member States, except that, for devices for self-testing, the instructions for use and the label must include a translation into the official language(
- s)of the Member State in which the device for self testing reaches its final user. 8.2. Where appropriate, the information to be supplied should take the form of symbols. Any symbol and identification colour used must conform to the harmonised standards. In areas for which no standards exist, the symbols and colour used must be described in the documentation supplied with the device. 8.3. In the case of devices containing a preparation which may be considered as being dangerous, taking account of the nature and quantity of its constituents and the form under which they are present, relevant danger symbols and labelling requirements of Directive 671548/EEC * and Directive 88/379/EEC † shall apply. Where there is insufficient space to put all the information on the device itself or on its label, the relevant danger symbols shall be put on the label and the other information required by those Directives shall be given in the instructions for use. *Council Directive 67/548/EEC of 27 June 1967 on the approximation of laws, regulations and administrative provisions relating to the classification, packaging and labeling of dangerous substances (OJ No. L 196, 16.08.67, P. 1 ). Directive as last amended by Commission Directive 97/69/EC (OJ No. L 343, 13.12.97, p. 19). †Council Directive 88/379/EEC of 7 June 1988 on the approximation of the laws, regulations and administrative provisions of the Member States relating to the classification, packaging and labelling of dangerous preparations (OJ No. L 187, 16.07.88, P. 14). Directive as last amended by Commission Directive 96/65/EC (OJ No. L 265, 18.10.96 p. 15). 20 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES The provisions of the aforementioned Directives on the safety data sheet shall apply, unless all relevant information as appropriate is already made available by the instructions for use. 8.4. The label must bear the following particulars which may take the form of symbols as appropriate: (
- a)the name or trade name and address of the manufacturer. For devices imported into the Community with a view to their distribution in the Community, the label, the outer packaging, or the instructions for use shall contain in addition the name and address of the authorised representative of the manufacturer; (
- b)the details strictly necessary for the user to uniquely identify the device and the contents of the packaging; (
- c)where appropriate, the word "STERILE" or a statement indicating any special microbiological state or state of cleanliness; (
- d)the batch code, preceded by the word "LOT", or the serial number; (
- e)if necessary, an indication of the date by which the device or part of it should be used, in safety, without degradation of performance, expressed as the year, the month and, where relevant, the day, in that order; (
- f)in case of devices for performance evaluation, the words "for performance evaluation only"; (
- g)where appropriate, a statement indicating the in vitro use of the device; (
- h)any particular storage and, or handling conditions; (
- i)where applicable, any particular operating instructions; (
- j)appropriate warnings and, or precautions to take; (
- k)if the device is intended for self-testing, that fact must be clearly stated. 8.5. If the intended purpose of the device is not obvious to the user, the manufacturer must clearly state the intended purpose in the instructions for use and, if appropriate, on the label. 8.6. Wherever reasonable and practicable, the devices and separate components must be identified, where appropriate in terms of batches, to allow all appropriate action to detect any potential risk posed by the devices and detachable components. 8.7. Where appropriate, the instructions for use must contain the following particulars: (
- a)the details referred to in point 8.4 of this section with the exception of points (
- d)and (e); (
- b)composition of the reagent product by nature and amount or concentration of the active ingredient(
- s)of the reagent(
- s)or kit as well as a statement, where appropriate, that the device contains other ingredients which might influence the measurement; (
- c)the storage conditions and shelf life following the first opening of the primary container, together with the storage conditions and stability of working reagents; (
- d)the performances referred to in point 3 of part A of this Schedule; (
- e)an indication of any special equipment required including information necessary for the identification of that special equipment for proper use; IN VITRO DIAGNOSTIC MEDICAL DEVICES (
- f)[ S.L.427.16 21 the type of specimen to be used, any special conditions of collection, pre-treatment and, if necessary, storage conditions and instructions for the preparation of the patient; (
- g)a detailed description of the procedure to be followed in using the device; (
- h)the measurement procedure to be followed with the device including as appropriate: - - the principle of the method, the specific analytical performance characteristics (e.g. sensitivity, specificity, accuracy, repeatability, reproducibility, limits of detection and measurement range, including information needed for the control of known relevant interferences), limitations of the method and information about the use of available reference measurement procedures and materials by the user, the details of any further procedure or handling needed before the device can be used (for example, reconstitution, incubation, dilution, instrument checks, etc.), the indication whether any particular training is required; (
- i)the mathematical approach upon which the calculation of the analytical result is made; (
- j)measures to be taken in the event of changes in the analytical performance of the device; (
- k)information appropriate to users on: (
- l)internal quality control including specific validation procedures, the traceability of the calibration of the device; the reference intervals for the quantities being determined, including a description of the appropriate reference population; (
- m)if the device must be used in combination with or installed with or connected to other devices or equipment in order to operate as required for its intended purpose, sufficient details of its characteristics to identify the correct devices or equipment to use in order to obtain a safe and proper combination; (
- n)all the information needed to verify whether the device is properly installed and can operate correctly and safely, plus details of the nature and frequency of the maintenance and calibration needed to ensure that the device operates properly and safely; information about safe waste disposal; (
- o)details of any further treatment or handling needed before the device can be used (for example, sterilisation, final assembly, etc.); (
- p)the necessary instructions in the event of damage to the protective packaging and details of appropriate methods of resterilisation or decontamination; (
- q)if the device is reusable, information on the appropriate processes to allow reuse, including cleaning, disinfection, packaging and resterilisation or decontamination, and any restriction on the number of reuses; (
- r)precautions to be taken as regards exposure, in reasonably foreseeable 22 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES environmental conditions, to magnetic fields, external electrical influences, electrostatic discharge, pressure or variations in pressure, acceleration, thermal ignition sources, etc.; (
- s)precautions to be taken against any special, unusual risks related to the use or disposal of the device including special protective measures; where the device includes substances of human or animal origin, attention must be drawn to their potential infectious nature; (
- t)specifications for devices for self-testing: - - - the results need to be expressed and presented in a way that is readily understood by a lay person; information needs to be provided with a device to the user on action to be taken (in case of positive, negative or indeterminate result) and on the possibility of false positive or false negative result, specific particulars may be omitted provided that the other information supplied by the manufacturer is sufficient to enable the user to use the device and to understand the result(
- s)produced by the device, the information provided must include a statement clearly directing that the user should not take any decision of medical relevance without first consulting his or her medical practitioner, the information must also specify that when the device for selftesting is used for the monitoring of an existing disease, the patient should only adapt the treatment if he has received the appropriate training to do so; (
- u)date of issue or latest revision of the instructions for use. Amended by: L.N. 371 of 2012. SCHEDULE II Based on Annex II of the Directive LIST OF DEVICES REFERRED TO IN REGULATIONS 7.2 AND 7.3 1. 2. List A - Reagents and reagent products, including related calibrators and control materials, for determining the following blood groups: ABO system, rhesus (C, c, D, E,
- e)anti-Kell, - reagents and reagent products, including related calibrators and control materials, for the detection, confirmation and quantification in human specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C and D, - Variant Creutzfeldt-Jakob disease (vCJD) assays for blood screening, diagnosis and confirmation. List B - Reagents and reagent products, including related calibrators and control materials, for determining the following blood groups: anti-Duffy and anti-Kidd, - reagents and reagent products, including related calibrators and control IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 23 materials, for determining irregular anti-erythrocytic antibodies, - reagents and reagent products, including related calibrators and control materials, for the detection and quantification in human samples of the following congenital infections: rubella, toxoplasmosis, - reagents and reagent products, including related calibrators and control materials, for diagnosing the following hereditary disease: phenylketonuria, - reagents and reagent products, including related calibrators and control materials, for determining the following human infections: cytomegalovirus, chlamydia, - reagents and reagent products, including related calibrators and control materials, for determining the following HLA tissue groups: DR, A, B, - reagents and reagent products, including related calibrators and control materials, for determining the following tumoral marker: PSA, - reagents and reagent products, including related calibrators, control materials and software, designed specifically for evaluating the risk of trisomy 21, - the following device for self-diagnosis, including its related calibrators and control materials: device for the measurement of blood sugar. SCHEDULE III Based on Annex III of the Directive EC DECLARATION OF CONFORMITY 1. The EC declaration of conformity is the procedure whereby the manufacturer or his authorised representative who fulfils the obligations imposed by sections 2 to 5 of this Schedule and additionally, in the case of devices for selftesting, the obligations imposed by section 6 of this Schedule, ensures and declares that the products concerned meet the provisions of these regulations which apply to them. The manufacturer must affix the CE marking in accordance with regulation 12 of these regulations. 2. The manufacturer must prepare the technical documentation described in section 3 of this Schedule and ensure that the manufacturing process follows the principles of quality assurance as set out in section 4 of this Schedule. 3. The technical documentation must allow assessment of the conformity of the product with the requirements of these regulations. It must include in particular: - a general description of the product, including any variants planned, - the documentation of the quality system, - design information, including the determination of the characteristics of the basic materials, characteristics and limitation of the performance of the devices, methods of manufacture and, in the case of instruments, design drawings, diagrams of components, sub-assemblies, circuits, etc., - in the case of devices containing tissues of human origin or substances derived from such tissue, information on the origin of such material and on the conditions in which it was collected, 24 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES - the descriptions and explanations necessary to understand the abovementioned characteristics, drawings and diagrams and the operation of the product, - the results of the risk analysis and, where appropriate, a list of the standards referred to in Article 5 of the Directive, applied in full or in part, and descriptions of the solutions adopted to meet the essential requirements of the Directive if the standards referred to in Article 5 of the Directive have not been applied in full, - in the case of sterile products or products with a special microbiological state or state of cleanliness, a description of the procedures used, - the results of the design calculations and of the inspections carried out, etc., - if the device is to be combined with other medical device(
- s)in order to operate as intended, proof must be provided that it conforms to the essential requirements when combined with any such device(
- s)having the characteristics specified by the manufacturer, - the test reports, - adequate performance evaluation data showing the performances claimed by the manufacturer and supported by a reference measurement system (when available), with information on the reference methods, the reference materials, the known reference values, the accuracy and measurement units used; such data should originate from studies in a clinical or other appropriate environment or result from relevant biographical references, - the labels and instructions for use, - the results of stability studies. 4. The manufacturer shall take necessary measures to ensure that the manufacturing process follows the principles of quality assurance as appropriate for the products manufactured. The system shall address: - the organisational structure and responsibilities, - the manufacturing processes and systematic quality control of production, - the means to monitor the performance of the quality system. 5. The manufacturer shall institute and keep up to date a systematic procedure to review experience gained from devices in the post-production phase and to implement appropriate means to apply any necessary corrective actions, taking account of the nature and risks in relation to the product. He shall notify the competent authorities of the following incidents immediately on learning of them: (
- i)any malfunction, failure or deterioration in the characteristics and, or performance of a device, as well as any inadequacy in the labelling or the instructions for use which, directly or indirectly, might lead to, or might have led to, the death of a patient or user or other persons or to a serious deterioration in his or their state of health; (
- ii)any technical or medical reason connected with the characteristics or the performance of a device for the reasons referred to in paragraph (
- i)leading to systematic recall of devices of the same type by the manufacturer. IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 25 6. For devices for self-testing the manufacturer shall lodge an application or examination of the medical design with a notified body. 6.1. The application shall enable the design of the device to be understood and shall enable conformity with the design-related requirements of the regulations to be assessed. It shall include: - test reports including, where appropriate, results of studies carried out with lay persons, - data showing the handling suitability of the device in view of its intended purpose for self-testing, - the information to be provided with the device on its label and its instructions for use. 6.2. The notified body shall examine the application and, if the design conforms to the relevant provisions of these regulations, shall issue the applicant with an EC design-examination certificate. The notified body may require the application to be completed by further tests or proof to allow assessment of conformity with the design-related requirements of the regulations. The certificate shall contain the conclusions of the examination, the conditions of validity, the data needed for identification of the approved design and, where appropriate, a description of the intended purpose of the product. 6.3. The applicant shall inform the notified body which issued the EC designexamination certificate of any significant change made to the approved design. Changes to the approved design must receive further approval from the notified body which issued the EC design-examination certificate wherever the changes could affect conformity with the essential requirements of the regulations or with the conditions prescribed for use of the product. This additional approval shall take the form of a supplement to the EC design-examination certificate. SCHEDULE IV Based on Annex IV of the Directive EC DECLARATION OF CONFORMITY (FULL QUALITY ASSURANCE SYSTEM) 1. The manufacturer must ensure application of the quality system approved for the design, manufacture and final inspection of the devices concerned, as specified in section 3 of this Schedule, and is subject to audit as laid down in point 3.3 and to the surveillance as specified in section 5 of this Schedule. In addition, the manufacturer must follow, for devices covered by Annex II, List A, of the Directive, which is set out in Schedule II, the procedures laid down in sections 4 and 6 of this Schedule. 2. The declaration of conformity is the procedure whereby the manufacturer who fulfils the obligations imposed by section 1 ensures and declares that the devices concerned meet the provisions of these regulations which apply to them. The manufacturer shall affix the CE marking in accordance with regulation 12 of these regulations, and shall draw up a declaration of conformity covering the devices concerned. 26 3. [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES Quality system 3.1. The manufacturer must lodge an application for assessment of his quality system with a notified body. The application must include: - the name and address of the manufacturer and any additional manufacturing site covered by the quality system, - adequate information on the device or device category covered by the procedure, - a written declaration that no such application has been lodged with any other notified body for the same device-related quality system, - the documentation on the quality system, - an undertaking by the manufacturer to fulfil the obligations imposed by the quality system approved, - an undertaking by the manufacturer to keep the approved quality system adequate and efficacious, - an undertaking by the manufacturer to institute and keep up to date a systematic procedure to review experience gained from devices in the post-production phase and to implement appropriate means to apply any necessary corrective action and notification as referred to in Annex III, section 5 of the Directive, which is set out in Schedule III. 3.2. Application of the quality system must ensure that the devices conform to the provisions of these regulations which apply to them at every stage, from design to final inspection. All the elements, requirements and provisions adopted by the manufacturer for his quality system must be documented in a systematic and orderly manner in the form of written policies and procedures, such as quality programmes, quality plans, quality manuals and quality records. It shall include in particular an adequate description of: (
- a)the manufacturer’s quality objectives; (
- b)the organisation of the business and in particular: - the organisational structures, the responsibilities of the managerial staff and their organisational authority where quality of design and manufacture of the devices is concerned, the methods of monitoring the efficient operation of the quality system and in particular its ability to achieve the desired quality of design and of product, including control of devices which fail to conform; (
- c)the procedures for monitoring and verifying the design of the devices and in particular: - a general description of the device, including any variants planned, all documentation referred to in Annex III, section 3, indents 3 to 13 of the Directive, which is set out in Schedule III, in the case of devices for self-testing, the information referred to in Annex III, section 6.1 of theþirective, which is set out in Schedule III, the techniques used to control and verify the design and the processes and systematic measures which will be used when the IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 27 devices are being designed; (
- d)the inspection and quality assurance techniques at the manufacturing stage and in particular: - the processes and procedures which will be used, particularly as regards sterilisation, the procedures in relation to purchasing, the product identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture; (
- e)the appropriate tests and trials which will be carried out before, during and after manufacture, the frequency with which they will take place, and the test equipment used; it must be possible to trace back the calibration. The manufacturer shall carry out the required controls and tests according to the latest state of the art. The controls and tests shall cover the manufacturing process including the characterisation of the raw material and the individual devices or each batch of devices manufactured. In testing the devices covered by Annex II, List A of the Directive, which is set out in Schedule II, the manufacturer shall take into account the most recent available information, in particular as regards the biological complexity and variability of the specimens to be tested with the in vitro device concerned. 3.3. The notified body must audit the quality system to determine whether it meets the requirements referred to in point 3.2 of this Schedule. It must presume that quality systems which implement the relevant harmonised standards conform to the requirements. The assessment team must have experience of assessments of the technology concerned. The assessment procedure must include an inspection on the manufacturer’s premises and, in duly substantiated cases, on the premises of the manufacturer ’s suppliers and, or subcontractors to inspect the manufacturing processes. The decision shall be notified to the manufacturer. It must contain the conclusions of the inspection and a reasoned assessment. 3.4. The manufacturer must inform the notified body which approved the quality system of any plan for substantial changes to the quality system or the product-range covered. The notified body must assess the changes proposed and verify whether after these changes the quality system still meets the requirements referred to in point 3.2 of this Schedule. It must notify the manufacturer of its decision. This decision must contain the conclusions of the inspection and a reasoned assessment. 4. Examination of the design of the product 4.1. For devices covered by Annex II, List A of the Directive, which is set out in Schedule II, in addition to the obligations imposed by section 3 of this Schedule, the manufacturer must lodge with the notified body an application for examination of the design dossier relating to the device which he plans to manufacture and which falls into the category referred to in point 3.1 of this Schedule. 4.2. The application must describe the design, manufacture and performances of the device in question. It must include the documents needed to assess whether the device conforms to the requirements of these regulations, as referred to in point 28 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES 3.2(
- c)of this Schedule. 4.3. The notified body must examine the application and, if the device conforms to the relevant provisions of the regulations, issue the application with an EC designexamination certificate. The notified body may require the application to be completed by further tests or proof to allow assessment of conformity with the requirements of the regulations. The certificate must contain the conclusions of the examination, the conditions of validity, the data needed for the identification of the approved design and, where appropriate, a description of the intended purpose of the device. 4.4. Changes to the approved design must receive further approval from the notified body which issued the EC design-examination certificate wherever the changes could affect conformity with the essential requirements of the regulations or with the conditions prescribed for use of the device. The applicant shall inform the notified body which issued the EC design-examination certificate of any such changes made to the approved design. The additional approval must take the form of a supplement to the EC design-examination certificate. 4.5. The manufacturer shall inform the notified body without delay if it has obtained information about changes to the pathogen and markers of infections to be tested, in particular as a consequence of biological complexity and variability. In this connection, the manufacturer shall inform the notified body whether any such change is likely to affect the performance of the in vitro diagnostic device concerned. 5. Surveillance 5.1. The aim of surveillance is to ensure that the manufacturer duly fulfils the obligations imposed by the approved quality system. 5.2. The manufacturer must authorise the notified body to carry out all the necessary inspections and supply it with all relevant information, in particular: - the documentation on the quality system, - the data stipulated in the part of the quality system relating to design, such as the results of analyses, calculation, tests, etc., - the data stipulated in the part of the quality system relating to manufacture, such as inspection reports and test data, calibration data, qualification reports of the personnel concerned, etc. 5.3. The notified body must periodically carry out appropriate inspections and assessments to make sure that the manufacturer applies the approved quality system and must supply the manufacturer with an assessment report. 5.4. In addition, the notified body may pay unannounced visits to the manufacturer. At the time of such visits, the notified body may, where necessary, carry out or ask for tests in order to check that the quality system is working properly. It must provide the manufacturer with an inspection report and, if a test has been carried out, with a test report. 6. Verification of manufactured products covered by Schedule II, List A 6.1. In the case of devices covered by Annex II, List A of the Directive, which is set out in Schedule II, the manufacturer shall forward to the notified body without delay after the conclusion of the controls and tests the relevant reports on the tests carried out on the manufactured devices or each batch of devices. Furthermore, the manufacturer shall make the samples of manufactured devices or batches of devices available to the notified body in accordance with pre- IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 29 agreed conditions and modalities. 6.2. The manufacturer may place the devices on the market, unless the notified body communicates to the manufacturer within the agreed time-frame, but not later than thirty days after reception of the samples, any other decision, including in particular any condition of validity of delivered certificates. SCHEDULE V Based on Annex V of the Directive EC TYPE-EXAMINATION 1. EC type-examination is the part of the procedure whereby a notified body ascertains and certifies that a representative sample of the production envisaged fulfils the relevant provisions of these regulations. 2. The application for EC type-examination shall be lodged by the manufacturer or by his authorised representative with a notified body. The application shall include: - the name and the address of the manufacturer and the name and address of the authorised representative if the application is lodged by the representative, - the documentation described in section 3 of this Schedule needed to assess the conformity of the representative sample of the production in question, hereinafter referred to as the "type", with the requirements of these regulations. The applicant shall make a "type" available to the notified body. The notified body may request other samples as necessary, - a written declaration that no application has been lodged with any other notified body for the same type. 3. The documentation must allow an understanding of the design, the manufacture and the performances of the device. The documentation shall contain the following items in particular: 4. - a general description of the type, including any variants planned, - all documentation referred to in Annex III, section 3, indents 3 to 13 of the Directive, which is set out in Schedule III, - in the case of devices for self-testing, the information referred to in Annex III, section 6.1 of the Directive, which is set out it Schedule III. The notified body shall: 4.1. examine and assess the documentation and verify that the type has been manufactured in conformity with that documentation; it shall also record the items designed in conformity with the applicable provisions of the standards referred to in Article 5 of the Directive, as well as the items not designed on the basis of the relevant provisions of the abovementioned standards; 4.2. perform or have performed appropriate examinations and the tests necessary to verify whether the solutions adopted by the manufacturer meet the essential requirements of these regulations if the standards 30 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES referred to in Article 5 of the Directive have not been applied; if the device is to be combined with other device(
- s)in order to operate as intended, proof must be provided that it conforms to the essential requirements when combined with any such device(
- s)having the characteristics specified by the manufactucer; 4.3. carry out or ask for the appropriate examinations and the tests necessary to verify whether, if the manufacturer has chosen to apply the relevant standards, these have actually been applied; 4.4. agree with the applicant on the place where the necessary examinations and tests will be carried out. 5. If the type conforms to the provisions of these regulations, the notified body shall issue the applicant with an EC type-examination certificate. The certificate shall contain the name and address of the manufacturer, the conclusions of the examination, the conditions of validity and the data needed for identification of the type approved. The relevant parts of the documentation shall be annexed to the certificate and a copy shall be kept by the notified body. 6. The manufacturer shall inform the notified body without delay if it has obtained information about changes to the pathogen and markers of infections to be tested, in particular as a consequence of biological complexity and variability. In this connection, the manufacturer shall inform the notified body whether any such change is likely to affect the performance of the in vitro device concerned. 6.1. Changes to the approved device must receive further approval from the notified body which issued the EC type-examination certificate wherever the changes may affect conformity with the essential requirements of the regulations or with the conditions prescribed for use of the device. The applicant shall inform the notified body which issued the EC type-examination certificate of any such change made to the approved device. This new approval shall take the form of a supplement to the initial EC type-exanunation certificate. 7. Administrative provisions 7.1. Other notified bodies may obtain a copy of the EC type-examination certificates and, or the supplements thereto. The annexes to the certificates must be available to the other notified bodies on reasoned application, after the manufacturer has been informed. SCHEDULE VI Based on Annex VI of the Directive EC VERIFICATION 1. EC verification is the procedure whereby the manufacturer or his authorized representative ensures and declares that the products which have been subject to the procedure set out in section 4 of this Schedule conform to the type described in the EC type-examination certificate and meet the requirements of these regulations which apply to them. 2.1. The manufacturer must take all the measures necessary to ensure that the manufacturing process produces products which conform to the type described in the EC type-examination certificate and the requirements of the regulations which apply to them. Before the start of manufacture, the manufacturer must prepare documents IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 31 defining the manufacturing process, in particular as regards sterilisation and the suitability of starting materials, where necessary, and define the necessary testing procedures according to the state of the art. All the routine, pre-established provisions must be implemented to ensure homogeneous production and conformity of the products with the type described in the EC type-examination certificate and with the requirements of these regulations which apply to them. 2.2. To the extent that for certain aspects the final testing according to point 6.3 of this Schedule is not appropriate, adequate process testing, monitoring and control methods shall be established by the manufacturer with the approval of the notified body. The provisions of Annex IV, section 5 of the Directive, which is set out in Schedule IV, shall apply accordingly in relation to the abovementioned approved procedures. 3. The manufacturer must undertake to institute and keep up to date a systematic procedure to review experience gained from devices in the postproduction phase and to implement appropriate means to apply any necessary corrective and notification action as referred to in Annex III, section 5 of the Directive, which is set out in Schedule III. 4. The notified body must carry out the appropriate examinations and tests taking account of point 2.2 of this Schedule in order to verify the conformity of the product with the requirements of the regulations either by examining and testing every product as specified in section 5 of this Schedule or by examining and testing products on a statistical basis as specified in section 6 of this Schedule, as the manufacturer decides. When carrying out statistical verification according to section 6, the notified body has to decide when statistical procedures for lot-by-lot inspection or isolated lot inspection have to be applied. Such decision must be taken in consultation with the manufacturer. In as far as the conduct of examinations and tests on a statistical basis is not appropriate, examinations and tests may be carried out on a random basis provided that such procedure in conjunction with the measures taken in accordance with point 2.2 of this Schedule ensures an equivalent level of conformity. 5. Verification by examination and testing of every product 5.1. Every product is examined individually and the appropriate tests defined in the relevant standard(
- s)referred to in Article 5 of the Directive or equivalent tests must be carried out in order to verify the conformity of the products with the EC type described in the type-examination certificate and with the requirements of the regulations which apply to them. 5.2. The notified body must affix, or have affixed, its identification number to each approved product and must draw up a written certificate of conformity relating to the tests carried out. 6. Statistical verification 6.1. The manufacturer must present the manufactured products in the form of homogeneous batches. 6.2. One or more random samples, as necessary, are taken from each batch. The products which make up the sample are examined and the appropriate tests defined in the relevant standard(
- s)referred to in Article 5 of the Directive or equivalent tests must be carried out to verify, where appropriate, the conformity of the products with the type described in the EC type-examination certificate and with the requirements of the regulations which apply to them in order to determine whether to accept or reject the batch. 32 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES 6.3. Statistical control of products will be based on attributes and, or variables, entailing sampling schemes with operational characteristics which ensure a high level of safety and performance according to the state of the art. The sampling scheme will be established by the harmonised standards referred to in Artic1e 5 of the Directive taking account of the specific nature of the product categories in question. 6.4. If the batch is accepted, the notified body affixes, or has affixed its identification number to each product and draws up a written certificate of conformity relating to the tests carried out. All products in the batch may be put on the market except any in the sample which failed to conform. If the batch is rejected the competent notified body must take appropriate measures to prevent the batch from being placed on the market. In the event of frequent rejection of batches, the notified body may suspend the statistical verification. The manufacturer may, on the responsibility of the notified body, affix the notified body’s identification number during the manufacturing process. SCHEDULE VII Based on Annex VII of the Directive EC DECLARATION OF CONFORMITY (PRODUCTION QUALITY ASSURANCE) 1. The manufacturer must ensure application of the quality system approved for the manufacture of the devices concerned and carry out the final inspection, as specified in section 3 of this Schedule, and is subject to the surveillance referred to in section 4 of this Schedule. 2. The declaration of conformity is the part of the procedure whereby the manufacturer who fulfils the obligations imposed by section 1 ensures and declares that the products concerned conform to the type described in the EC typeexanunation certificate and meet the provisions of these regulations which apply to them. The manufacturer must affix the CE marking in accordance with regulation 12 of these regulations and draw up a declaration of conformity covering the devices concerned. 3. Quality system 3.1. The manufacturer must lodge an application for assessment of his quality system with a notified body. The application must include the technical documentation on the types approved and a copy of the EC type-examination certificates. 3.2. Application of the quality system must ensure that the devices conform to the type described in the EC type-examination certificate. All the elements, requirements and provisions adopted by the manufacturer for his quality system must be documented in a systematic and orderly manner in the form of written policy statements and procedures. This quality system documentation must permit uniform interpretation of the quality policy and procedures such as quality progranunes, plans, manuals and records. IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 33 It must include in particular an adequate description of: (
- a)the manufacturer’s quality objectives; (
- b)the organisation of the business and in particular: - the organisational structures, the responsibilities of the managerial staff and their organisational authority where quality of manufacture of the devices is concerned, the methods of monitoring the efficient operation of the quality system and in particular its ability to achieve the desired quality of product, including control of devices which fail to conform; (
- c)the inspection and quality assurance techniques at the manufacturing stage and in particular: - the processes and procedures which will be used, particularly as regards sterilisation, the procedures in relation to purchasing, the product identification procedures drawn up and kept up to date from drawings, specifications or other relevant documents at every stage of manufacture; (
- d)the appropriate tests and trials to be carried out before, during and after manufacture, the frequency with which they will take place, and the test equipment used; it must be possible to trace back the calibration. 3.3. The notified body must audit the quality system to determine whether it meets the requirements referred to in section 3.2 of this Schedule. It must presume that quality systems which implement the relevant harmonised standards conform to these requirements. The assessment team must have past experience of assessments of the technology concerned. The assessment procedure must include an inspection on the manufacturer’s premises and, in duly substantiated cases, on the premises of the manufacturer ’s suppliers and, or subcontractors to inspect the manufacturing processes. The decision must be notified to the manufacturer. It must contain the conclusions of the inspection and a reasoned assessment. 3.4. The manufacturer shall inform the notified body which approved the quality system of any plan for substantial changes to the quality system. The notified body must assess the changes proposed and verify whether after these changes the quality system still meets the requirements referred to in section 3.2 of this Schedule. It must notify the manufacturer of its decision. This decision must contain the conclusions of the inspection and a reasoned assessment. 4. Surveillance The provisions of Annex IV, section 5 of the Directive, which is set out in Schedule IV, shall apply. 5. Verification of manufactured products covered by Schedule II, List A. 5.1. In the case of devices covered by Annex II, List A of the Directive, which is set out in Schedule II, the manufacturer shall forward to the notified body without delay after the conclusion of the controls and tests the relevant reports on the tests carried out on the manufactured devices or each batch of devices. Furthermore, the manufacturer shall make the samples of manufactured devices or batches of devices available to the notified body in accordance with pre- 34 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES agreed conditions and modalities. 5.2. The manufacturer may place the devices on the market, unless the notified body communicates to the manufacturer within the agreed time-frame, but not later than thirty days after reception of the samples, any other decision, including in particular any condition of validity of delivered certificates. SCHEDULE VIII Based on Annex VIII of the Directive STATEMENT AND PROCEDURES CONCERNING DEVICES FOR PERFORMANCE EVALUATION 1. For devices for performance evaluation the manufacturer or his authorized representative shall draw up the statement containing the information stipulated in section 2 of this Schedule and ensure that the relevant provisions of these regulations are met. 2. The statement shall contain the following information: - data allowing identification of the device in question, - an evaluation plan stating in particular the purpose, scientific, technical or medical grounds, scope of the evaluation and number of devices concerned, - the list of laboratories or other institutions taking part in the evaluation study, - the starting date and scheduled duration for the evaluations and, in the case of devices for self-testing, the location and number of lay persons involved, - a statement that the device in question conforms to the requirements of the regulations, apart from the aspects covered by the evaluation and apart from those specifically itenused in the statement, and that every precaution has been taken to protect the health and safety of the patient, user and other persons. 3. The manufacturer shall also undertake to keep available for the competent national authorities the documentation allowing an understanding of the design, manufacture and performances of the product, including the expected performances, so as to allow assessment of conformity with the requirements of these regulations. This documentation must be kept for a period ending at least five years after the end of the performance evaluation. The manufacturer shall take all the measures necessary for the manufacturing process to ensure that the products manufactured conform to the documentation mentioned in the first paragraph. 4. The provisions of regulations 9.1, 9.3 and 9.5 of these regulations shall apply to devices intended for performance evaluation. IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 35 SCHEDULE IX Based on Annex IX of the Directive CRITERIA FOR THE DESIGNATION OF NOTIFIED BODIES 1. The notified body, its director and the assessment and verification staff shall not be the designer, manufacturer, supplier, installer or user of the devices which they inspect, nor the authorised representative of any of these persons. They may not be directly involved in the design, construction, marketing or maintenance of the devices, nor represent the parties engaged in these activities. This in no way precludes the possibility of exchanges of technical information between the manufacturer and the body. 2. The notified body and its staff must carry out the assessment and verification operations with the highest degree of professional integrity and the requisite competence in the field of medical devices and must be free from all pressures and inducements, particularly financial, which might influence their judgment or the results of the inspection, especially from persons or groups of persons with an interest in the results of the verifications. Should the notified body subcontract specific tasks connected with the establishment and verification of the facts, it must first ensure that the subcontractor meets the provisions of the regulations. The notified body shall keep at the disposal of the national authorities the relevant documents assessing the subcontractor’s qualifications and the work carried out by the subcontractor under these regulations. 3. The notified body must be able to carry out all the tasks assigned to such bodies by one of Annexes III to VII of the Directive, which are set out in Schedules III to VII, and for which it has been notified, whether these tasks are carried out by the body itself or on its responsibility. In particular, it must have the necessary staff and possess the facilities needed to perform properly the technical and administrative tasks entailed in assessment and verification. This includes the availability of sufficient scientific staff within the organisation who possess adequate experience and knowledge necessary to assess the biological and medical functionality and performance of devices for which it has been notified, in relation to the requirements of this Directive and, in particular, with the requirements of Annex I of the Directive, which is set out in Schedule I. The notified body must also have access to the equipment necessary for the verifications required. 4. The inspection staff must have: - sound vocational training covering all the assessment and verification operations for which the body has been designated, - satisfactory knowledge of the rules on the inspections which they carry out and adequate experience of such inspections, - the ability required to draw up the certificates, records and reports to demonstrate that the inspections have been carried out. 5. The impartiality of the inspection staff must be guaranteed. Their remuneration must not depend on the number of inspections carried out, nor on the results of the inspections. 6. The body must take out civil liability insurance, unless liability is assumed by the State under domestic legislation or the Member State itself carries out the inspections directly. 7. The staff of the inspection body are bound to observe professional secrecy with regard to all information gained in the course of their duties (except vis-a-vis 36 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES the competent administrative authorities of the State in which their activities are carried out) under these regulations or any provision of national law putting it into effect. SCHEDULE X Based on Annex X of the Directive CE MARKING OF CONFORMITY The CE conformity marking shall consist of the initials "CE" taking the following form: For the avoidance of doubt, it is hereby decdared that the grid providing the background in the above graduated drawing is not part of the CE marking. - if the marking is reduced or enlarged the proportions given in the above graduated drawing must be respected, - the various components of the CE marking must have substantially the same vertical dimension, which may not be less than 5mm. This minimum dimension may be waived for small-scale devices. Added by: L.N. 30 of 2004. Substituted by: L.N. 21 of 2010. SCHEDULE XI Based on Annex I of Commission Decision 2009/108/EC of 3 February 2009 on common technical specifications (CTS) for in vitro diagnostic medical devices 1. SCOPE The common technical specifications set out in this Schedule shall apply for the purposes of Schedule II, List A. 2. DEFINITIONS AND TERMS IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 37 (Diagnostic) sensitivity The probability that the device gives a positive result in the presence of the target marker. True positive A specimen known to be positive for the target marker and correctly classified by the device. False negative A specimen known to be positive for the target marker and misclassified by the device. (Diagnostic) specificity The probability that the device gives a negative result in the absence of the target marker. False positive A specimen known to be negative for the target marker and misclassified by the device. True negative A specimen known to be negative for the target marker and correctly classified by the device. Analytical sensitivity Analytical sensitivity may be expressed as the limit of detection, i.e. the smallest amount of the target marker that can be precisely detected. Analytical specificity Analytical specificity means the ability of the method to determine solely the target marker. Nucleic acid amplification techniques (NAT) The term ‘NAT’ is used for tests for the detection and/or quantification of nucleic acids by either amplification of a target sequence, by amplification of a signal or by hybridisation. Rapid test ‘Rapid test’ means qualitative or semi quantitative in vitro diagnostic medical devices, used singly or in a small series, which involve non-automated procedures and have been designed to give a fast result. Robustness The robustness of an analytical procedure means the capacity of an analytical procedure to remain unaffected by small but deliberate variations in method parameters and provides an indication of its reliability during normal usage. Whole system failure rate The whole system failure rate means the frequency of failures when entire process is performed as prescribed by the manufacturer. the Confirmation assay Confirmation assay means an assay used for the confirmation of a reactive result from a screening assay. Virus typing assay Virus typing assay means an assay used for typing with already known positive samples, not used for primary diagnosis of infection or for screening. Sero-conversion HIV samples 38 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES Sero-conversion HIV samples mean: - p24 antigen and/or HIV RNA positive and; - recognised by all of the antibody screening tests and; - positive or indeterminate confirmatory assays. Early sero-conversion HIV samples Early sero-conversion HIV samples mean: - p24 antigen and/or HIV RNA positive and; - not recognised by all of the antibody screening tests and; - indeterminate or negative confirmatory assays. 3. COMMON TECHNICAL SPECIFICATIONS (CTS) FOR PRODUCTS REFERRED TO IN SCHEDULE II, LIST A. 3.1 CTS for performance evaluation of reagents and reagent products for the detection, confirmation and quantification in human specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D: General Principles: 3.1.1 Devices which detect virus infections placed on the market for use as either screening or diagnostic tests, shall meet the requirements for sensitivity and specificity set out in Table 1. See also principle 3.1.11 for screening assays. 3.1.2 Devices intended by the manufacturer for testing body fluids other than serum or plasma, for example urine, saliva, etc shall meet the same CTS requirements for sensitivity and specificity as serum or plasma tests. The performance evaluation shall test samples from the same individuals in both the tests to be approved and in a respective serum or plasma assay. 3.1.3 Devices intended by the manufacturer for self-test, i.e. home use, shall meet the same CTS requirements for sensitivity and specificity as respective devices for professional use. Relevant parts of the performance evaluation shall be carried out (or repeated) by appropriate lay users to validate the operation of the device and the instructions for use. 3.1.4 All performance evaluations shall be carried out in direct comparison with an established state of the art device. The device used for comparison shall be one bearing CE marking, if on the market at the time of the performance evaluation. 3.1.5 If discrepant test results are identified as part of an evaluation, these results shall be resolved as far as possible, for example: - by evaluation of the discrepant sample in further test systems; - by use of an alternative method or marker; - by a review of the clinical status and diagnosis of the patient; and - by the testing of follow-up-samples. 3.1.6 Performance evaluations shall equivalent to the European population. be performed on a population 3.1.7 Positive specimens used in the performance evaluation shall be selected to reflect different stages of the respective disease(s), different antibody patterns, different genotypes, different subtypes, mutants, etc. 3.1.8 Sensitivity with true positives and sero-conversion samples shall be IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 39 evaluated as follows: 3.1.8.1 Diagnostic test sensitivity during sero-conversion has to represent the state of the art. Whether further testing of the same or additional sero-conversion panels is conducted by the notified body or by the manufacturer the results shall confirm the initial performance evaluation data (see Table 1). Seroconversion panels should start with a negative bleed(
- s)and should have narrow bleeding intervals. 3.1.8.2 For blood screening devices (with the exception of HBsAg and antiHBc tests), all true positive samples shall be identified as positive by the device to be CE marked (Table 1). For HBsAg and anti-HBc tests the new device shall have an overall performance at least equivalent to that of the established device (see 3.1.4). 3.1.8.3 Regarding HIV tests: - all sero-conversion HIV samples shall be identified as positive; and - at least 40 early sero-conversion HIV samples shall be tested. Results should conform to the state of the art. 3.1.9 Performance evaluation of screening assays shall include 25 positive (if available in the case of rare infections) "same day" fresh serum and/or plasma samples (≤ 1 day after sampling). 3.1.10 Negative specimens used in a performance evaluation shall be defined so as to reflect the target population for which the test is intended, for example blood donors, hospitalised patients, pregnant women, etc. 3.1.11 For performance evaluations for screening assays (Table 1) blood donor populations shall be investigated from at least two blood donation centres and consist of consecutive blood donations, which have not been selected to exclude first time donors. 3.1.12 Devices shall have a specificity of at least 99.5% on blood donations, unless otherwise indicated in the accompanying tables. Specificity shall be calculated using the frequency of repeatedly reactive (i.e. false positive) results in blood donors negative for the target marker. 3.1.13 Devices shall be evaluated to establish the effect of potential interfering substances, as part of the performance evaluation. The potential interfering substances to be evaluated will depend to some extent on the composition of the reagent and configuration of the assay. Potential interfering substances shall be identified as part of the risk analysis required by the essential requirements for each new device but may include, for example: - specimens representing "related" infections; - specimens from multipara, i.e. women who have had more than one pregnancy, or rheumatoid factor positive patients; - for recombinant antigens, human antibodies to components of the expression system, for example anti E. coli, or anti yeast. 3.1.14 For devices intended by the manufacturer to be used with serum and plasma the performance evaluation must demonstrate serum to plasma equivalency. This shall be demonstrated for at least 50 donations (25 positive and 25 negative). 3.1.15 For devices intended for use with plasma the performance evaluation shall verify the performance of the device using all anticoagulants which the manufacturer indicates for use with the device. This shall be demonstrated 40 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES for at least 50 donations (25 positive and 25 negative). 3.1.16 As part of the required risk analysis the whole system failure rate leading to false-negative results shall be determined in repeat assays on lowpositive specimens. 3.1.17 If a new in vitro diagnostic medical device belonging to Schedule II, List A is not specifically covered by the common technical specification, the common technical specification for a related device should be taken into account. Related devices may be identified on different grounds, for example by the same or similar intended use or by similar risks. 3.2 Additional Requirements for HIV antibody/antigen combined tests 3.2.1 HIV antibody/antigen combined tests intended for anti-HIV and p24 antigen detection which include claims for single p24 antigen detection shall follow Table 1 and Table 5, including criteria for analytical sensitivity for p24 antigen. 3.2.2 HIV antibody/antigen combined tests intended for anti-HIV and p24 detection which do not include claims for single p24 detection shall follow Table 1 and Table 5, excluding criteria for analytical sensitivity for p24. 3.3 Additional Requirements for Nucleic Acid Amplification Techniques (NAT) The performance evaluation criteria for NAT assays can be found in Table 2. 3.3.1 For target sequence amplification assays, a functionality control for each test sample (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection. 3.3.2 The analytical sensitivity or detection limit for NAT assays shall be expressed by the 95% positive cut-off value. This is the analyte concentration where 95% of test runs give positive results following serial dilutions of an international reference material for example a WHO standard or calibrated reference material. 3.3.3 Genotype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped samples. 3.3.4 Results of quantitative NAT assays shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application. 3.3.5 NAT assays may be used to detect virus in antibody negative samples, i.e. pre sero conversion samples. Viruses within immune-complexes may behave differently in comparison to free viruses, for example during a centrifugation step. It is therefore important that during robustness studies, antibody-negative (pre sero-conversion) samples are included. 3.3.6 For investigation of potential carry-over, at least five runs with alternating high positive and negative specimens shall be performed during robustness studies. The high positive samples shall comprise of samples with naturally occurring high virus titres. 3.3.7 The whole system failure rate leading to false-negative results shall be determined by testing low-positive specimens. Low positive specimens shall contain a virus concentration equivalent to 3 times the 95% positive cut-off virus concentration. 3.4 CTS for the manufacturer’s release testing of reagents and eagent products IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 41 for the detection, confirmation and quantification in human specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D (Immunological assays only). 3.4.1 The manufacturer’s release testing criteria shall ensure that every batch consistently identifies the relevant antigens, epitopes, and antibodies. 3.4.2. The manufacturer’s batch release testing for screening assays shall include at least 100 specimens negative for the relevant analyte. 3.5 CTS for performance evaluation of reagents and reagent products for determining the following blood group antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K). Criteria for performance evaluation of reagents and reagent products for determining the blood groups antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group B 9 system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K) can be found in Table 9. 3.5.1 All performance evaluations shall be carried out in direct comparison with an established state of the art device. The device used for comparison shall be one bearing CE marking, if on the market at the time of the performance evaluation. 3.5.2 If discrepant test results are identified as part of an evaluation, these results shall be resolved as far as possible, for example: - by evaluation of the discrepant sample in further test systems; - by use of an alternative method. 3.5.3 Performance evaluations shall equivalent to the European population. be performed on a population 3.5.4 Positive specimens used in the performance evaluation shall be selected to reflect variant and weak antigen expression. 3.5.5 Devices shall be evaluated to establish the effect of potential interfering substances, as part of the performance evaluation. The potential interfering substances to be evaluated will depend to some extent on the composition of the reagent and configuration of the assay. Potential interfering substances shall be identified as part of the risk analysis required by the essential requirements for each new device. 3.5.6 For devices intended for use with plasma the performance evaluation shall verify the performance of the device using all anticoagulants which the manufacturer indicates for use with the device. This shall be demonstrated for at least 50 donations. 3.6. CTS for the manufacturers release testing of reagents and reagent products for determining the blood group antigens: ABO blood group system ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K). 3.6.1 The manufacturer’s release testing criteria shall ensure that every batch consistently identifies the relevant antigens, epitopes, and antibodies. 3.6.2 Requirements for manufacturers batch release testing are outlined in Table 10. Table 1: "Screening" assays: anti-HIV 1 and 2, anti-HTLV I and II, 42 [ S.L.427.16 IN VITRO DIAGNOSTIC MEDICAL DEVICES anti-HCV, HBsAg, anti-HBc anti-HIV-1/2 Diagnostic sensitivity Positive specimens 400 HIV-1 100 HIV-2 including 40 non-Bsubtypes, all available HIV/ 1 subtypes should be represented by at least 3 samples per subtype anti-HTLVanti-HCV I/II 300 HTLV-I 400 (positive 100 HTLV-II samples) Including samples from different stages of infection and reflecting different antibody patterns. HBsAg anti-HBc 400 Including subtypeconsideration 400 Including evaluation of other HBVmarkers To be defined when available Genotype 1-4: > 20 samples - per genotype (including non-a sub-types of genotype 4); 5: > 5 samples; Seroconversion panels Analytical sensitivity Standards Specificity Unselected donors (including1 st time donors) Hospitalize d patients Potentially crossreacting bloodspecimens (RF+, related viruses, pregnant women, etc) 20 panels 10 further panels (at Notified Body or manufacturer) To be defined when available 6: if available 20 panels 10 further panels (at Notified Body or manufacturer) 5000 5000 5000 20 panels 10 further panels (at Notified Body or manufacturer) 0.130 IU/ml (Second International Standard for HBsAg, subtype adw2, genotype A, NIBSC code: 00/588) 5000 200 200 200 200 200 100 100 100 100 100 5000 IN VITRO DIAGNOSTIC MEDICAL DEVICES [ S.L.427.16 43 Table 2: NAT assays for HIV1, HCV, HBV, HTLV I/II (qualitative and quantitative; not molecular typing) HIV1 NAT qualitative HCV quantitative qualitative HBV quantitative quantitative As for HIV quantitative As for HIV quantitative HTLV I/II qualitative quantitative As for HIV quantitative Sensitivity Detection limit Detection of analytical sensitivity (IU/ml; defined on WHO standards or calibrated reference materials) According to EP validation guideline